Cedex® Bio HT Analyzer
Powerful And Comprehensive Cell Culture Profiling
Cedex Bio HT Analyzer - the right instrument for improved workflow efficiency
- High throughput: Numerous samples can be processed simultaneously with minimal manual intervention
- Long-term operation: Long time on-board stability of reagents and assay cal
- Flexible dilution: Automated sample dilution capabilities save time and accuracy
- Full automation: Online integration with automated systems to optimize measurement points and timelines
- Enhanced flexibility with the 96-well upgrade kit
Application
Use the Cedex Bio HT Analyzer for fast and reliable analysis of your bioprocess.
The instrument delivers precise results for > 30 parameters of high importance for bio-process control in process development as well as in manufacturing.
Properties
The technology of the Cedex Bio Analyzer is based on photometric analysis.
Test Throughput (Theoretical Max):
- 150-300 tests/hour
Sample Types:
- Mammalian cell culture
- Microbial fermentation
Sample Handling:
- Up to 90 sample tubes (six 15-sample racks). If using 96-well plate + two 15-sample racks, up to 126 samples
- Cooled rack position for controls and calibrators
- Automated sample dilution
- Barcode reading via laser scanner, with immediate STAT recognition
Sample Container Types:
- Sampling tubes: 5-10 mL; 16 x 100; 16 x 75; 13 x 100; 13 x 75 mm
- Standard cup: 1.5 mL Micro
- Micro cup: 650 μL
- Cup on tube definable
Sample Volume:
- Minimal sample volume of 100 μL for several tests
- 2-20 μL sample consumption per test
Sample Barcode Types:
- Code 128, Codabar, Code 2 of 5 interleaved, code 39
Reagents Onboard Capacity:
- Up to 32 Cedex Bio HT reagent cassettes in 8 racks on-board
- 50-800 tests per cassette
- Cooled reagent compartment (10-15°C), on-board stability up to 3 months
- Calibration typically once per lot
Reaction Cells:
- Holds 1,000 disposable cuvettes with 5 mm path length and 120-240 μL reaction volume
Specification
Dimensions
- 135.0 cm x 66.0 cm x 75.0 cm (W x D x H)
Weight
- Packed: 315 kg (694 lb)
- Unpacked (instrument only): 184 kg (406 lb)
Energy Requirement
- 100-125/200-240 volts AC, 50 or 60 Hz, consumption 1200 VA
Control Unit
- HP workstation running Windows 10, Intel Core i5 with 8 GB RAM
- Dual 500 GB hard drives, DVDRW drive, Ethernet
- Weight: 5 kg (11 lb)
Accesory box
- Weight: 10.5 kg (231 lb)
System Interfaces
- RS232 serial interface for HOST connection, Ethernet for remote access
Water Requirements
- Up to 2 liters per hour in operation mode, Type 1 NCCLS
Operating Conditions
- Ambient temperature: +15 to +32°C
- Ambient humidity: 30 to 80% (RH, without condensation)
Certification
- CE, UL, C-UL
Regulatory Disclaimer
For use in quality control/manufacturing process only.